Connection with | PC USB and Bluetooth connection |
Power Supply Usb | DC 5v or two AA batteries |
Power Consumption | < 200mA from USB or batteries |
Reports | Summary with target, max, table, narrative, DTS... Rep beats by ead, selected strips, user selections, integrated Stress ECG and Echo report. |
Protocols | Industry leading number of pre-programmed protocols, treadmill, ergometer, pharmacologic, custom, post test review |
Interpretation | NAI comprehensive interpretation program |
Leads | Standard 12-lead |
Lead Leakage | < 10uA |
Acquisition Mode | Simultaneous interpretive 12-lead acquisition |
Sampling Rate | 12-lead sampling rate 250-1000Hz |
A/D Rate | 1kHz |
Time Base | 12.5, 25 and 50 mm/sec |
Dynamic Range | ± 5 mV |
Frequency Response | 0.05 to 150 Hz |
Gains | 2.5, 5, 10, 10/5, 20, 40 mm/mV |
Input Impedance | >1012 ohms |
System Noise | <40 uV RTI |
ECG Interpretation Classes | Mls, blocks, enlargements, axis and AFib |
Filters | AC, baseline, muscle artifact |
CMRR | >120 dB |
Networking | Built-in |
Measurements | HR with global measurements, complete interval values and ST information |
Dimensions | 2.75" W x 1.25" D x 4.87" H (6.985 cm x 3.175 cm x 12.364 cm) |
Weight | 0.26 lb (0.12 kg) |
PC Requirements | Tablet/Laptop/Desktop PC, Intel Compatible, 4GB RAM, 40GB HD, WINDOWS XP through WINDOWS 7, 8, 10 and MAC's with Boot Camp, Dual Boot,VMware Fusion or Parallels |